| Автор | I G Donaldson |
| Дата выпуска | 1967-12-01 |
| dc.description | A theory has been developed for a microflowmeter of the type described by Benseman. This theory has been found to match the performance of an experimental model of this type and has been used to forecast that instruments of this design could be capable of measuring flows in the range of 0-5 × 10<sup>-3</sup> g cm<sup>-2</sup> sec<sup>-1</sup> with a sensitivity of 0.2 μg cm<sup>-2</sup> sec<sup>-1</sup> per microvolt output. |
| Формат | application.pdf |
| Издатель | Institute of Physics Publishing |
| Название | Theoretical consideration of a microflowmeter |
| Тип | paper |
| DOI | 10.1088/0950-7671/44/12/313 |
| Print ISSN | 0950-7671 |
| Журнал | Journal of Scientific Instruments |
| Том | 44 |
| Первая страница | 1018 |
| Последняя страница | 1022 |
| Аффилиация | I G Donaldson; Physics and Engineering Laboratory, Department of Scientific and Industrial Research, Lower Hutt, New Zealand |
| Выпуск | 12 |