Автор C E Bassey
Автор O O Ojo
Автор I Akpabio
Дата выпуска 1991-09-01
dc.description The repeat profile of an X-ray department in a developing country was analysed monthly between July 1989 and June 1990. Results showed an average repeat rate of 3.7% for the period of study. The main causes of film repetition were: equipment fault, 33.9%; radiographer's fault, 27.4%; film fault, 19.3%; processing fault, 10.8% and patient's fault, 8.6%. The average repeat rate in the first six months of study reduced by 50% in the last six months. This was due to the effectiveness of implementation of corrective actions. The overall repeat rate was found to correlate well with both the equipment fault and radiographer's fault with correlation coefficients, r, of 0.94 and 0.91, respectively. It is expected that a further reduction in the repeat rate will be obtained after the introduction of quality assurance programmes.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Repeat profile analysis in an x-ray department
Тип paper
DOI 10.1088/0952-4746/11/3/003
Electronic ISSN 1361-6498
Print ISSN 0952-4746
Журнал Journal of Radiological Protection
Том 11
Первая страница 179
Последняя страница 183
Аффилиация C E Bassey; Ilorin Univ., Nigeria
Аффилиация O O Ojo; Ilorin Univ., Nigeria
Аффилиация I Akpabio; Ilorin Univ., Nigeria