An improved MOSFET-based Robinson oscillator for NMR detection
K J Wilson; C P G Vallabhan; K J Wilson; Dept. of Phys., Cochin Univ. of Sci. & Technol., India; C P G Vallabhan; Dept. of Phys., Cochin Univ. of Sci. & Technol., India
Журнал:
Measurement Science and Technology
Дата:
1990-05-01
Аннотация:
The design and performance of a modified version of the Robinson oscillator using MOSFETs are described. The NMR detector, which offers excellent sensitivity, is capable of operating at very low RF levels (down to 1 mV) and permits satisfactory recording of weak NMR signals even at room temperatures. The circuit is very simple to implement and requires no critical components or layout.