Дата выпуска 1996-11-01
dc.description A novel algorithm for the reconstruction of dynamic images using diametric excitation has been developed. The algorithm is specifically designed to image impedance changes in the brain using boundary data obtained from scalp electrodes by incorporating a priori information. The a priori information is obtain by solving the forward problem using a finite-element model (FEM) which includes the discontinuity of the skull resistivity. The advantages with this new approach are that the sensitivity and accuracy of the location of the impedance changes are improved compared to methods based on adjacent excitation.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Improvement of the positional accuracy of EIT images of the head using a Lagrange multiplier reconstruction algorithm with diametric excitation
Тип paper
DOI 10.1088/0967-3334/17/4A/008
Electronic ISSN 1361-6579
Print ISSN 0967-3334
Журнал Physiological Measurement
Том 17
Первая страница A49
Последняя страница A57
Выпуск 4A