Автор |
A L Vitouchkine |
Автор |
J E Faller |
Дата выпуска |
2002-10-01 |
dc.description |
We describe the design of and first results from a cam-driven portable absolute gravimeter. The cambased design of a dropper allows us to build a compact system that uses 20 mm of free fall for the measurement. The instrument drops a test mass 200 times a minute (a measurement rate of more than 3 times per second). The centre of mass of the apparatus is kept fixed during the entire measurement cycle, thus avoiding ``recoil'' effects that could systematically bias the measurement. Seismic isolation is provided by a passive eddy-current-damped simple spring. Measurement results taken at Table Mountain Gravity Observatory (TMGO, Boulder, Colorado), though reproducible at a level of several microgals (1 µGal = 10<sup>-8</sup> m/s<sup>2</sup>), show a sensitivity to the measurement start and stop points at the 20 µGal level. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Measurement results with a small cam-driven absolute gravimeter |
Тип |
paper |
DOI |
10.1088/0026-1394/39/5/6 |
Electronic ISSN |
1681-7575 |
Print ISSN |
0026-1394 |
Журнал |
Metrologia |
Том |
39 |
Первая страница |
465 |
Последняя страница |
469 |
Аффилиация |
A L Vitouchkine; Joint Institute for Laboratory Astrophysics (JILA), National Institute of Standards and Technology and University of Colorado, Boulder, CO 80309-0440, USA |
Аффилиация |
J E Faller; Joint Institute for Laboratory Astrophysics (JILA), National Institute of Standards and Technology and University of Colorado, Boulder, CO 80309-0440, USA |
Выпуск |
5 |