Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор G J Budgell
Автор C Martens
Автор F Claus
Дата выпуска 2001-11-01
dc.description The delivery efficiency of step and shoot intensity modulated radiotherapy (IMRT) has been improved by the installation of fast-tuning magnetrons into three travelling wave linear accelerators. The IMRT delivery efficiency and the beam start-up performance have been compared before and after installation. Start-up and inter sub-field times were reduced by an average of 3.0 s. A typical start-up time from depression of the start button to beam on is now around 4 s. Delivery efficiency for a variety of clinical and quality control prescriptions was improved by an average of 30.7% (range 7.4–60.9%), depending on a complex combination of the number of sub-fields, distance moved by leaves and dose rate. For the oldest accelerator (7 years old), dosimetric accuracy was significantly improved for low dose sub-fields. The dose output was within 2% for a 1 monitor unit (MU) sub-field and 1% for a 2 MU sub-field. The two newer accelerators displayed similar or better dose characteristics even before fast-tuning magnetron installation. Beam symmetries and flatnesses were acceptable at all energies and dose rates, and showed no obvious degradation in low dose sub-fields. It is recommended that fast-tuning magnetrons are adopted for accelerators of this design performing step and shoot IMRT.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Improved delivery efficiency for step and shoot intensity modulated radiotherapy using a fast-tuning magnetron
Тип note
DOI 10.1088/0031-9155/46/11/402
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 46
Первая страница N253
Последняя страница N261
Выпуск 11

Скрыть метаданые