The delivery efficiency of step and shoot intensity modulated radiotherapy (IMRT) has been improved by the installation of fast-tuning magnetrons into three travelling wave linear accelerators. The IMRT delivery efficiency and the beam start-up performance have been compared before and after installation. Start-up and inter sub-field times were reduced by an average of 3.0 s. A typical start-up time from depression of the start button to beam on is now around 4 s. Delivery efficiency for a variety of clinical and quality control prescriptions was improved by an average of 30.7% (range 7.4–60.9%), depending on a complex combination of the number of sub-fields, distance moved by leaves and dose rate. For the oldest accelerator (7 years old), dosimetric accuracy was significantly improved for low dose sub-fields. The dose output was within 2% for a 1 monitor unit (MU) sub-field and 1% for a 2 MU sub-field. The two newer accelerators displayed similar or better dose characteristics even before fast-tuning magnetron installation. Beam symmetries and flatnesses were acceptable at all energies and dose rates, and showed no obvious degradation in low dose sub-fields. It is recommended that fast-tuning magnetrons are adopted for accelerators of this design performing step and shoot IMRT. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Improved delivery efficiency for step and shoot intensity modulated radiotherapy using a fast-tuning magnetron |
Тип |
note |
DOI |
10.1088/0031-9155/46/11/402 |
Electronic ISSN |
1361-6560 |
Print ISSN |
0031-9155 |
Журнал |
Physics in Medicine and Biology |
Том |
46 |
Первая страница |
N253 |
Последняя страница |
N261 |
Выпуск |
11 |