Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор J Pérez-Calatayud
Автор F Lliso
Автор F Ballester
Автор M A Serrano
Автор J L Lluch
Автор Y Limami
Автор V Puchades
Автор E Casal
Дата выпуска 2001-07-01
dc.description The CSM3 <sup>137</sup>Cs type stainless-steel encapsulated source is widely used in manually afterloaded low dose rate brachytherapy. A specially asymmetric source, CSM3-a, has been designed by CIS Bio International (France) substituting the eyelet side seed with an inactive material in the CSM3 source. This modification has been done in order to allow a uniform dose level over the upper vaginal surface when this `linear' source is inserted at the top of the dome vaginal applicators. In this study the Monte Carlo GEANT3 simulation code, incorporating the source geometry in detail, was used to investigate the dosimetric characteristics of this special CSM3-a <sup>137</sup>Cs brachytherapy source. The absolute dose rate distribution in water around this source was calculated and is presented in the form of an along-away table. Comparison of Sievert integral type calculations with Monte Carlo results are discussed.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A Monte Carlo study of dose rate distribution around the specially asymmetric CSM3-a <sup>137</sup>Cs source
Тип note
DOI 10.1088/0031-9155/46/7/401
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 46
Первая страница N169
Последняя страница N174
Выпуск 7

Скрыть метаданые