Автор | Kenneth E Ekstrand |
Автор | William H Hinson |
Автор | J Daniel Bourland |
Автор | Allan F deGuzman |
Автор | Volker W Stieber |
Автор | Steven B Tatter |
Автор | Thomas L Ellis |
Дата выпуска | 2003-12-21 |
dc.description | We have investigated the use of an adapter that permits the use of a Leksell coordinate frame with a linear accelerator stereotactic radiosurgery system based on the Brown–Robert–Wells (BRW) design. This device is useful when lesions that are planned for treatment on a Leksell Gamma Knife system are found to be inaccessible to the Gamma Knife. We have found that with this device objects within a head phantom can be targeted by the linear accelerator within an accuracy of approximately 1 mm. |
Формат | application.pdf |
Издатель | Institute of Physics Publishing |
Копирайт | 2003 IOP Publishing Ltd |
Название | The use of a Leksell-BRW adapter for linac radiosurgery as an adjunct to Gamma Knife treatment |
Тип | paper |
DOI | 10.1088/0031-9155/48/24/008 |
Electronic ISSN | 1361-6560 |
Print ISSN | 0031-9155 |
Журнал | Physics in Medicine and Biology |
Том | 48 |
Первая страница | 4105 |
Последняя страница | 4110 |
Выпуск | 24 |