Автор Kenneth E Ekstrand
Автор William H Hinson
Автор J Daniel Bourland
Автор Allan F deGuzman
Автор Volker W Stieber
Автор Steven B Tatter
Автор Thomas L Ellis
Дата выпуска 2003-12-21
dc.description We have investigated the use of an adapter that permits the use of a Leksell coordinate frame with a linear accelerator stereotactic radiosurgery system based on the Brown–Robert–Wells (BRW) design. This device is useful when lesions that are planned for treatment on a Leksell Gamma Knife system are found to be inaccessible to the Gamma Knife. We have found that with this device objects within a head phantom can be targeted by the linear accelerator within an accuracy of approximately 1 mm.
Формат application.pdf
Издатель Institute of Physics Publishing
Копирайт 2003 IOP Publishing Ltd
Название The use of a Leksell-BRW adapter for linac radiosurgery as an adjunct to Gamma Knife treatment
Тип paper
DOI 10.1088/0031-9155/48/24/008
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 48
Первая страница 4105
Последняя страница 4110