Автор Maia, Ana F
Автор Caldas, Linda V E
Дата выпуска 2006-03-21
dc.description Tandem systems, each formed by a pencil ionization chamber with and without a specific covering, were developed and tested in standard radiation beams. These systems were designed to be used in computed tomography radiation beams, where the half-value layer (HVL) determination is not an easy task. The behaviour of the tandem systems in diagnostic radiology showed the possibility of their use to confirm HVL values previously determined by the conventional HVL measurement method in quality control programmes. These systems also have other advantages: low cost, easy application and quick measurement procedure.
Формат application.pdf
Издатель Institute of Physics Publishing
Копирайт 2006 IOP Publishing Ltd
Название A simple method for evaluation of half-value layer variation in CT equipment
Тип paper
DOI 10.1088/0031-9155/51/6/016
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 51
Первая страница 1595
Последняя страница 1601
Аффилиация Maia, Ana F; Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, Av. Prof. Lineu Prestes, 2242, CEP 05422-970, São Paulo, Brazil
Аффилиация Caldas, Linda V E; Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, Av. Prof. Lineu Prestes, 2242, CEP 05422-970, São Paulo, Brazil