The measurement of Pitot pressure in high enthalpy expansion tubes
M A Sutcliffe; R G Morgan; M A Sutcliffe; Center for Hypersonics, The University of Queensland, Brisbane 4072, Australia; R G Morgan; Center for Hypersonics, The University of Queensland, Brisbane 4072, Australia
Журнал:
Measurement Science and Technology
Дата:
2001-03-01
Аннотация:
Expansion tubes operating at total flow enthalpies of 100 MJ kg<sup>-1</sup> or more have characteristical test times of 30-50 µs. Under these conditions, the response time of the Pitot pressure measuring device is critical when performing flow calibration studies. The conventional technique of using a commercial pressure transducer protected by shielding has not always proven to be effective, due to the relatively large (and variable) response time caused by the shielding. A device called the stress wave bar gauge has been designed and calibrated and shown to be an effective way to measure the Pitot pressure with a response time of only 2-3 µs.