Автор Victor D Irby
Дата выпуска 2004-09-01
dc.description The concept and subsequent experimental verification of the proportionality between pulse amplitude and detector transit time for microchannel-plate detectors is presented. This discovery has led to considerable improvement in the overall timing resolution for detection of high-energy γ-photons. Utilizing a <sup>22</sup>Na positron source, a full width half maximum (FWHM) timing resolution of 138 ps has been achieved. This FWHM includes detector transit-time spread for both chevron-stack-type detectors, timing spread due to uncertainties in annihilation location, all electronic uncertainty and any remaining quantum mechanical uncertainty. The first measurement of the minimum quantum uncertainty in the time interval between detection of the two annihilation photons is reported. The experimental results give strong evidence against instantaneous spatial localization of γ-photons due to measurement-induced nonlocal quantum wavefunction collapse. The experimental results are also the first that imply momentum is conserved only after the quantum uncertainty in time has elapsed (Yukawa H 1935 Proc. Phys. Math. Soc. Japan 17 48).
Формат application.pdf
Издатель Institute of Physics Publishing
Копирайт 2004 IOP Publishing Ltd
Название Picosecond timing resolution detection of γ-photons utilizing microchannel-plate detectors: experimental tests of quantum nonlocality and photon localization
Тип paper
DOI 10.1088/0957-0233/15/9/018
Electronic ISSN 1361-6501
Print ISSN 0957-0233
Журнал Measurement Science and Technology
Том 15
Первая страница 1799
Последняя страница 1810
Аффилиация Victor D Irby; Department of Physics, University of South Alabama, Mobile, Al 36688-0002, USA
Выпуск 9

Скрыть метаданые