Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор Stefan Nikles
Автор Robert Bradley
Автор Sanford Bledsoe
Автор Khalil Najafi
Дата выпуска 2004-07-01
dc.description A study was conducted to measure and characterize the reliability of polysilicon cantilever beams with electroplated gold pads for use in a high-density biomedical connector. In this design, an array of beams is brought into contact with a corresponding array of 30 µm high gold bumps, forming electrical connection. Analytical computations of multi-layer beams were performed, including the effects of residual stresses. Beam leads of various lengths and widths with electroplated gold contact pads on their ends were tested over 1000 cycles to determine their mechanical reliability, and to measure their contact resistance with gold bumps on a separate substrate. The dimensions of the polysilicon beam that produce the least breakage were determined to be 400 µm long by 125 µm wide. For a beam having a calculated contact force of ∼100 µN, the initial contact resistance was 764 mΩ. After 1000 connect/disconnect cycles, beams of this type had an average final contact resistance of 1.598 Ω. These results demonstrate that very high-density connectors with high mechanical reliability and low-contact resistance can be fabricated.
Формат application.pdf
Издатель Institute of Physics Publishing
Копирайт 2004 IOP Publishing Ltd
Название Design and testing of conductive polysilicon beam leads for use in a high-density biomedical connector
Тип paper
DOI 10.1088/0960-1317/14/7/017
Electronic ISSN 1361-6439
Print ISSN 0960-1317
Журнал Journal of Micromechanics and Microengineering
Том 14
Первая страница 957
Последняя страница 968
Выпуск 7

Скрыть метаданые