Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор C L Giusca
Автор I Severn
Автор J Greenwood
Дата выпуска 2008-07-01
dc.description The effective area of a hydraulic pressure balance is typically characterised using the cross-floating technique, where the pressure balance is compared against a standard instrument. This method usually involves operating both instruments in the pressure transmitting fluid of the standard, even if the tested device is normally operated with a different fluid. We describe novel experiments to determine the influence of various commonly used oils on the measured effective area of the balances. Measurements were made using five different oils in the pressure range 28 MPa to 200 MPa. Our results show that in some cases the contribution to the effective area when using different oils was nearly 60 ppm. Although a difference in the effective area with respect to the liquid viscosity was intuitively expected, the results appear to show a greater dependency on the liquid density, which contradicts the current assumptions used to formulate the theory describing the pressure balance.
Формат application.pdf
Издатель Institute of Physics Publishing
Копирайт © 2008 IOP Publishing Ltd
Название The influence of hydraulic fluids on pressure balance calibration results
Тип paper
DOI 10.1088/1742-6596/121/15/152001
Electronic ISSN 1742-6596
Print ISSN 1742-6588
Журнал Journal of Physics: Conference Series
Том 121
Первая страница 152001
Последняя страница 152004
Аффилиация C L Giusca; National Physical Laboratory, Hampton Road, Teddington, UK, TW11 0LW
Аффилиация I Severn; National Physical Laboratory, Hampton Road, Teddington, UK, TW11 0LW
Аффилиация J Greenwood; National Physical Laboratory, Hampton Road, Teddington, UK, TW11 0LW
Выпуск 15

Скрыть метаданые