Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор V C Spanoudaki
Автор A B Mann
Автор A N Otte
Автор I Konorov
Автор I Torres-Espallardo
Автор S Paul
Автор S I Ziegler
Дата выпуска 2007-12-01
dc.description We propose in this study a novel PET detector concept as insert for simultaneous PET/MR imaging, using arrays of Silicon Photomultipliers (SiPMs) as photodetectors, read out by a data acquisition system based on sampling ADCs. A 2 × 2 LSO-SiPM detector array and four single channel LYSO-SiPM detectors have been evaluated and compared to a LSO-APD detector. A 17.9% energy resolution and a 1.4 ns time resolution have been measured. No degradation of these values could be detected when simultaneous MR acquisitions were performed. The non-linear detector behaviour due to the limited dynamic range and recovery time effects has been studied. In addition, the contribution of dark counts and optical crosstalk for PET applications was also addressed. The feasibility for position localization of the incident light to a SiPM array using Anger logic has been investigated.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Use of single photon counting detector arrays in combined PET/MR: Characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector
Тип paper
DOI 10.1088/1748-0221/2/12/P12002
Electronic ISSN 1748-0221
Журнал Journal of Instrumentation
Том 2
Первая страница P12002
Последняя страница P12002
Выпуск 12