Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор Diethrich, Edward, B
Автор Bahadir, Ilhan
Автор Mandile, Gary
Автор Gordon, Michael
Дата выпуска 1987
dc.description A new ventricular assist device (VAD) developed by Symbion, Inc., received its initial biventricular clinical use in a 53-year-old man suffering cardiogenic shock from severe congestive heart failure secondary to cardiomyopathy and a 15-year history of coronary artery disease.The eight-day period of biventricular support on the Symbion devices was satisfactory in spite of two complications probably unrelated to the pump. First, a neurologic deficit was noted after transition from the temporary support pumps to the Symbion pumps, ostensibly from a small air bolus entering the aortic return line. The deficit was deemed mild and reversible, and the patient showed improvement. Secondly, poor renal function secondary to prolonged hypotension associated with cardiac arrest persisted throughout the support period. All other organs, however, were well supported.On the eighth day of support, a dramatic change in hepatic function and neurologic deterioration justified termination of support. At explant of the devices, a kink and an occluding thrombus were found at the inferior port of the right atrial outflow cannula, undoubtedly the cause of the declining hepatic status. Clots were also found adherent to both VAD chambers at the air vents, but no evidence of emboli was found at autopsy.Redesign of the air vent sites and cannulae may be required to improve the Symbion VAD's performance.
Издатель Sage Publications
Название A new ventricu lar assist device for acute cardiac failure: report of initial use for biventricular support
Тип Other
DOI 10.1177/026765918700200403
Print ISSN 0267-6591
Журнал Perfusion
Том 2
Первая страница 245
Последняя страница 261
Аффилиация Diethrich, Edward, B, Departments of Cardiovascular Surgery and Cardiology, Arizona Heart Institute, Phoenix
Аффилиация Bahadir, Ilhan, Departments of Cardiovascular Surgery and Cardiology, Arizona Heart Institute, Phoenix
Аффилиация Mandile, Gary, Departments of Cardiovascular Surgery and Cardiology, Arizona Heart Institute, Phoenix
Аффилиация Gordon, Michael, Departments of Cardiovascular Surgery and Cardiology, Arizona Heart Institute, Phoenix
Выпуск 4
Библиографическая ссылка Beckman D., Siderys H.Prolonged left heart bypass with the BioMedicus vortex centrifugal pump. Contemp Surg1986; 29: 69-73.
Библиографическая ссылка Rose DM, Laschinger J., Grossi E. et al. Experimental and clinical results with a simplified left heart assist device for treatment of profound left ventricular dysfunction. World J Surg 1985; 9: 11-17.
Библиографическая ссылка Pae WE, Pierce WS, Pennock JL, Campbell DB, Waldhausen JALongterm results of ventricular assist pumping in postcardiotomy cardiogenic shock. J Thorac Cardiovasc Surg1987; 93: 434-41.
Библиографическая ссылка Magovern GJ , Park SB, Maher TD Use of a centrifugal pump without anticoagulants for postoperative left ventricular assist. World J Surg 1985 ; 9: 25-36.
Библиографическая ссылка Pennington DG , Merjavy JP, Swartz MT et al. The importance of biventricular failure in patients with postoperative cardiogenic shock . Ann Thorac Surg 1985; 39: 16-26.
Библиографическая ссылка Pennington DG , Samuels LD, Williams G. et al. Experience with the Pierce-Donachy ventricular assist device in postcardiotomy patients with cardiogenic shock . World J Surg 1985; 9: 37-46.
Библиографическая ссылка Rose DM, Colvin SB, Culliford AT et al. Late functional and hemodynamic status of surviving patients following insertion of the left heart assist device. J Thorac Cardiovasc Surg 1983; 86: 639-45.
Библиографическая ссылка Park SB, Liebler GA, Burkholder JA et al. Mechanical support of the failing heart. Ann Thorac Surg 1986; 42: 627-31.
Библиографическая ссылка Pennington DG , Bernhard WF, Golding LR et al. Longterm follow-up of postcardiotomy patients with profound cardiogenic shock treated with ventricular assist devices . Circulation 1985; 72(Suppl II): 11216-26.
Библиографическая ссылка Pennock JL, Pierce WS, Wisman GB, Bull. AP, Waldhausen JASurvival and complications following ventricular assist pumping for cardiogenic shock. Ann Surg1983; 198: 469-78.
Библиографическая ссылка Pierce WS, Parr Gvs, Myers JL et al. Clinical effectiveness of mechanical ventricular bypass in treating postoperative heart failure. Artif Organs 1983; 7: 25-30.
Библиографическая ссылка Pierce WS, Parr Gvs, Myers JL et al. Ventricular assist pumping in patients with cardiogenic shock after cardiac operations. N. Engl J Med 1981 ; 305: 1606-10.
Библиографическая ссылка Gaines WE, Pierce WS, Donachy JH et al. The Pennsylvania State University paracorporeal ventricular assist pump: optimal methods of use. World J Surg 1985; 9: 47-53.
Библиографическая ссылка Richenbacher WE, Pierce WSRight ventricular failure following implantation of a left ventricular assist device. Curr Surg1983; 40: 274-77.
Библиографическая ссылка O'Neill MJ, Pierce WS, Wisman CB, Osbakken MD, Parr Gvs, Waldhausen JASuccessful management of right ventricular failure with the ventricular assist pump following aortic valve replacement and coronary bypass grafting . J Thorac Cardiovasc Surg 1984; 87: 106-11.
Библиографическая ссылка Pierce WSThe implantable ventricular assist device. J Thorac Cardiovasc Surg1984; 87; 811-13.
Библиографическая ссылка Farrar DJ, Compton PG, Hershon JJ, Hill JDRight ventricular function in an operating room model of mechanical left ventricular assistance and its effects in patients with depressed left ventricular function . Circulation 1985; 72: 1279-85.
Библиографическая ссылка Kolff J., Deeb GMArtificial heart and left ventricular assist devices. Surg Clin North Am1985; 65: 661-77.
Библиографическая ссылка Norman JC, Igo SRMechanical circulatory assistance: established (IABP) and evolving (LVAD). A narrative summary. Thorac Cardiovasc Surg1985; 33: 133-45.
Библиографическая ссылка Watson JTThe present and future of cardiac assist devices. Artif Organs1985; 9: 138-43.
Библиографическая ссылка Zumbro GL, Shearer G.Kitchens WR, Galloway RIMechanical assistance for biventricular failure following coronary artery bypass and heart transplantation. Heart Transplant1985; 4: 348-52.
Библиографическая ссылка Relman ASArtificial heart: permanent and temporary. N Engl J Med1986; 314: 644-45.
Библиографическая ссылка Hill JD, Farrar DJ, Hershon JJ et al. Use of a prosthetic ventricle as a bridge to cardiac transplantation for postinfarction cardiogenic shock. N Engl J Med 1986; 314: 626-28.
Библиографическая ссылка Carpentier A., Perier P., Brugger JP et al. Heterotopic artificial heart as a bridge to cardiac transplantation. Lancet 1986; ii: 97-98.
Библиографическая ссылка Pennock JL, Wisman CB, Pierce WSMechanical support of the circulation prior to cardiac transplantation . Heart Transplant 1982; 1: 299-305.
Библиографическая ссылка Pennington CG, Codd JE, Merjavy JP et al. The expanded use of ventricular bypass systems for severe cardiac failure and as a bridge to cardiac transplantation . Heart Transplant 1984; 3: 170-75.
Библиографическая ссылка Macgovern JA , Pennock JL, Campbell DB, Pae WE, Pierce WS, Waldhausen JA Bridge to heart transplantation. J. Heart Transplant 1986; 5: 196-202.
Библиографическая ссылка Joyce LD, Pritzker MR, Kiser JC et al. Use of the mini Jarvik-7 total artificial heart as a bridge to transplantation. J. Heart Transplant 1986 ; 5: 203-209.
Библиографическая ссылка Griffith BP , Kormos RL, Wei LM, Borovetz HS, Trento A., Hardesty RL Use of the total artifical heart as an interim device: initial experience in Pittsburgh with four patients. J. Heart Transplant 1986; 5: 210-14.
Библиографическая ссылка Joyce LD, Johnson KE, Pierce WS et al. Summary of the world experience with clinical use of total artificial hearts as heart support devices. J. Heart Transplant 1986; 5: 229-35.
Библиографическая ссылка Devries WC, Anderson JL, Joyce LD et al. Clinical use of the total artifical heart. N Engl J Med 1984; 310: 273-78.

Скрыть метаданые