Автор |
Alston, RP |
Автор |
Rennie, RA |
Автор |
Toal, F. |
Автор |
Kenny, Gnc |
Дата выпуска |
1989 |
dc.description |
A computerized system which incorporated haemodynamic, temperature and arterial pH (pHa) monitoring was developed to record data and to improve temperature and pHa control during hypothermic cardiopulmonary bypass (CPB). The system was evaluated in the course of a study of haemodynamics and metabolism during CPB. It was planned to have three 10 minute periods of study in 24 patients, during which time arterial and nasopharyngeal temperatures would be maintained between 27° and 29°C and pHa between 7.38 and 7.42. Repeated recordings of haemodynamic variables, temperature and pHa were made during each period. Only 59 of a possible 72 periods were obtained due to shortage of time at stable hypothermia. The mean number of total readings was 73.5 (70.1, 77.0) and the mean number of acceptable readings was 71.2 (67.3, 75.1). Good nasopharyngeal temperature control was achieved, with all the 95% confidence intervals of the means falling within the prescribed limits. Control of arterial temperature was less good as there were five periods in which the 95% confidence intervals of the means fell outwith the prescribed range. Poor pHa control was achieved since there were 16 periods in which the 95% confidence intervals of the mean were outside the preset range. The system functioned well in recording data and resulted in good temperature control, but poor pHa control. |
Издатель |
Sage Publications |
Название |
A computerized system forthe study of oxygen uptake and haemodynamics during cardiopulmonary bypass |
Тип |
Journal Article |
DOI |
10.1177/026765918900400305 |
Print ISSN |
0267-6591 |
Журнал |
Perfusion |
Том |
4 |
Первая страница |
197 |
Последняя страница |
203 |
Аффилиация |
Alston, RP, Department of Anaesthesia, University of Glasgow |
Аффилиация |
Rennie, RA, Department of Anaesthesia, University of Glasgow |
Аффилиация |
Toal, F., Department of Anaesthesia, University of Glasgow |
Аффилиация |
Kenny, Gnc, Department of Anaesthesia, University of Glasgow |
Выпуск |
3 |
Библиографическая ссылка |
Harris EA, Seelye ER, Squire AWOxygen consumption during cardiopulmonary bypass with moderate hypothermia . BrJAnesth 1971; 43: 1114-20. |
Библиографическая ссылка |
Fox LS, Blackstone EH, Kirklin JW, Stewart RW, Samuelson PNRelationship of whole body oxygen consumption to perfusion flow rate during hypothermic cardiopulmonary bypass. J Thorac Cardiovasc Surg1982; 83 : 239-48. |
Библиографическая ссылка |
Hickey PR, Buckley MJ, Philbin DMPulsatile and nonpulsatile bypass: review of a counterproductive controversy. Ann Thorac Surg1983; 36: 721-37. |
Библиографическая ссылка |
Harken AHHydrogen ion concentration and oxygen uptake in an isolated canine hind limb . J Appl Physiol 1976; 40: 1-5. |
Библиографическая ссылка |
Paterson RWEffect of PaCO2 on oxygen consumption during cardiopulmonary bypass in man. Anaesth Analg1976; 55: 269-73. |
Библиографическая ссылка |
Springer RR, Clark DK, Lea AS, Salis RTEffects of changes in carbon dioxide tension on oxygen consumption during cardiopulmonary bypass. Chest1979; 75: 549-54. |
Библиографическая ссылка |
Taylor KM, Maxted KJ, Bain WHIn vitro evaluation of the Stockert roller pump system. Proc Euro Soc Art Org1978; 4. |
Библиографическая ссылка |
Gourly T.Computers in perfusion practice. Perfusion1987; 2: 79-85. |
Библиографическая ссылка |
Vainionpää V. , Timisjarvi J. Spontaneous oscillation of the systemic arterial pressure during cardiopulmonary bypass in man. The effects of some drugs used. Basic Res Cardiol 1987; 82: 178-85. |
Библиографическая ссылка |
Alston RP, Trew A.An in vitro assessment of a monitor for continuous inline measurement of PO2, PCO2 and pH. Perfusion1987; 2: 139-47. |
Библиографическая ссылка |
Alston RPClinical evaluation of a monitor for inline measurement of PO 2, PCO2 and pH during cardiopulmonary bypass. Perfusion1988; 3 : 225-31. |
Библиографическая ссылка |
White FNHypothermia: lessons from comparative physiology. In: Utley JR ed, Pathophysiology and techniques of cardiopulmonary bypass , volume 1. London : Williams and Wilkins, 1982: 145-58. |
Библиографическая ссылка |
Cox DRPlanning of experiments. New York: Wiley, 1958: 91-153. |
Библиографическая ссылка |
Davis FM, Parimelzhagen KN, Harris EAThermal balance during cardiopulmonary bypass with moderate hypothermia in man. Br J Anaesth1977; 49: 1127-32. |
Библиографическая ссылка |
Kent B., Pearce ECOxygen consumption during cardiopulmonary bypass in the uniformly cool dog . J Appl Physiol 1974; 37: 917-22. |
Библиографическая ссылка |
Pearson DTBlood gas control during cardiopulmonary bypass. Perfusion1988; 3 : 113-33. |
Библиографическая ссылка |
Pearson DT, Waterhouse PS, Holden MP, Paslad SJ, Murray A.A clinical evaluation of the gas transfer characteristics and gaseous microemboli production of two bubble oxygenators. Life Support Syst1984; 2 : 252-66. |