We present a method for characterizing the adsorption of solutes in microfluidic devices that is sensitive to both long-lived and transient adsorption and can be applied to a variety of realistic device materials, designs, fabrication methods, and operational parameters. We have characterized the adsorption of two highly adsorbing molecules (FITC-labeled bovine serum albumin (BSA) and rhodamine B) and compared these results to two low adsorbing species of similar molecular weights (FITC-labeled dextran and fluorescein). We have also validated our method by demonstrating that two well-known non-fouling strategies [deposition of the polyethylene oxide (PEO)-like surface coating created by radio-frequency glow discharge plasma deposition (RF-GDPD) of tetraethylene glycol dimethyl ether (tetraglyme, CH3O(CH2CH2O)4CH3), and blocking with unlabeled BSA] eliminate the characteristic BSA adsorption behavior observed otherwise. |
Формат |
application.pdf |
Издатель |
Royal Society of Chemistry |
Название |
A method for characterizing adsorption of flowing solutes to microfluidic device surfaces |
Тип |
research-article |
DOI |
10.1039/b612894g |
Electronic ISSN |
1473-0189 |
Print ISSN |
1473-0197 |
Журнал |
Lab on a Chip |
Том |
7 |
Первая страница |
281 |
Последняя страница |
285 |
Аффилиация |
Hawkins Kenneth R.; Department of Bioengineering, University of Washington |
Аффилиация |
Steedman Mark R.; UCSF/UC Berkeley Joint Graduate Group in Bioengineering, University of California at San Francisco and Berkeley |
Аффилиация |
Baldwin Richard R.; Department of Bioengineering, University of Washington |
Аффилиация |
Fu Elain; Department of Bioengineering, University of Washington |
Аффилиация |
Ghosal Sandip; Department of Mechanical Engineering, Northwestern University |
Аффилиация |
Yager Paul; Department of Bioengineering, University of Washington |
Выпуск |
2 |
Библиографическая ссылка |
Nakanishi, J. Biosci. Bioeng., 2001, 91, 233 |
Библиографическая ссылка |
Lionello, Lab Chip, 2005, 5, 254 |
Библиографическая ссылка |
Linder, Anal. Chem., 2001, 73, 4181 |
Библиографическая ссылка |
Munson, Lab Chip, 2004, 4, 438 |
Библиографическая ссылка |
Ghosal, Anal. Chem., 2002, 74, 771 |
Библиографическая ссылка |
Jordan, Anal. Chem., 1997, 69, 1449 |
Библиографическая ссылка |
Popat, Biosens. Bioelectron., 2004, 19, 1037 |
Библиографическая ссылка |
Liu, J. Biomed. Mater. Res., 2002, 60, 126 |
Библиографическая ссылка |
Liu, Anal. Chem., 2000, 72, 5939 |
Библиографическая ссылка |
Shen, J. Biomater. Sci., Polym. Ed., 2001, 12, 961 |
Библиографическая ссылка |
Shen, Langmuir, 2003, 19, 1692 |
Библиографическая ссылка |
Zhang, Biotechnol. Bioeng., 1997, 56, 618 |
Библиографическая ссылка |
Schroen, Langmuir, 1995, 11, 3068 |
Библиографическая ссылка |
Bange, Biosens. Bioelectron., 2005, 20, 2488 |
Библиографическая ссылка |
Beyer, J. Biomed. Mater. Res., 1997, 36, 181 |
Библиографическая ссылка |
Hatch, Proc. IEEE, 2004, 92, 126 |
Библиографическая ссылка |
Hawkins, Lab Chip, 2003, 3, 248 |