Автор M O Norris
Автор J H Strange
Дата выпуска 1969-12-01
dc.description A cryostat is described which has been used for sample temperature control in a pulsed n.m.r. spectrometer in the range 78°K to room temperature. The temperature controller may be preset to the sample temperature required and the sample subsequently maintained at this temperature to better than 0·5 degK for long periods of time. The advantages of this method of temperature control over the more conventional gas-flow and controlled heat-leak methods are discussed.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A nuclear magnetic resonance sample temperature controller using liquid nitrogen injection
Тип note
DOI 10.1088/0022-3735/2/12/428
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 2
Первая страница 1106
Последняя страница 1108
Аффилиация M O Norris; The Physics Laboratory, The University, Canterbury, Kent
Аффилиация J H Strange; The Physics Laboratory, The University, Canterbury, Kent
Выпуск 12