Автор |
F D G Bennett |
Автор |
D G Thorpe |
Дата выпуска |
1970-03-01 |
dc.description |
Gain degradation measurements have been carried out on five 419BL and two 318BL Mullard lead-type channel electron multipliers, over periods up to several thousand hours, at a pressure of less than 10 ntorr. The devices showed reductions of about one order of magnitude on their initial gain values. The reduction occurred in 10 h at an output count rate of 2 × 10<sup>5</sup> counts s<sup>−1</sup> and several thousand hours at 2 × 10<sup>2</sup> counts s<sup>−1</sup>. At the high output count rate, the initial gain was found to be depressed by over an order of magnitude on the nominal value of 10<sup>8</sup>. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Gain degradation of lead-type channel electron multipliers in ultra-high vacuum |
Тип |
note |
DOI |
10.1088/0022-3735/3/3/420 |
Print ISSN |
0022-3735 |
Журнал |
Journal of Physics E: Scientific Instruments |
Том |
3 |
Первая страница |
241 |
Последняя страница |
243 |
Аффилиация |
F D G Bennett; SRC Radio and Space Research Station, Ditton Park, Slough, Bucks. |
Аффилиация |
D G Thorpe; SRC Radio and Space Research Station, Ditton Park, Slough, Bucks. |
Выпуск |
3 |