| Автор | R B Marcus | 
| Автор | T T Sheng | 
| Автор | C J Calbick | 
| Дата выпуска | 1972-03-01 | 
| dc.description | A high-vacuum specimen chamber was constructed for a Siemens electron microscope in order that in situ studies of the growth of refractory metal films may be performed under clean well-defined conditions. The chamber is differentially pumped and reaches a vacuum of 0.1 mu Torr with the rest of the column at about 20 mu Torr. The hydrocarbon contamination rate at the specimen is approximately 2.5 AA min<sup>-1</sup> compared with about 500 AA min<sup>-1</sup> for the unmodified instrument. The stage operates smoothly, and images at 80000* are steady and free from vibration. | 
| Формат | application.pdf | 
| Издатель | Institute of Physics Publishing | 
| Название | A high vacuum large volume specimen chamber for an electron microscope | 
| Тип | paper | 
| DOI | 10.1088/0022-3735/5/3/014 | 
| Print ISSN | 0022-3735 | 
| Журнал | Journal of Physics E: Scientific Instruments | 
| Том | 5 | 
| Первая страница | 234 | 
| Последняя страница | 236 | 
| Аффилиация | R B Marcus; Bell Telephone Labs. Inc., Murray Hill, NJ, USA | 
| Аффилиация | T T Sheng; Bell Telephone Labs. Inc., Murray Hill, NJ, USA | 
| Аффилиация | C J Calbick; Bell Telephone Labs. Inc., Murray Hill, NJ, USA | 
| Выпуск | 3 | 
 
 
 